Abstract

Background

Stereotactic biopsy is an important and minimally invasive technique used for a variety of indications in neurosurgery. Initially, this technique required a frame, but recently there have been a number of newer, less cumbersome approaches to biopsy including robotic arms, fixed arms, and, more recently, skull-mounted miniframes. Miniframes are attractive because they are disposable and low profile. However, the relatively limited degree of freedom offered by currently available devices necessitates a preplanned burr hole, which in turn limits flexibility and multiple trajectories. The AXiiiS device is a skull-mounted, magnetic resonance imaging-compatible miniframe that provides a similar degree of freedom with a frame while maintaining a low-profile, disposable platform.

Objective

To assess the image-guided trajectory alignment accuracy using AXiiiS stereotactic miniframe biopsy of intracranial lesions. The accuracy of the AXiiiS device is compared with the Navigus Trajectory Guide as platforms.

Methods

After approval by our institutional review board, medical records of 10 neurosurgical patients with intracranial pathologies were chosen for AXiiiS stereotactic miniframe biopsy, and histological correlation was obtained.

Results

Ten reported cases demonstrate the precision and ease of using the AXiiiS stereotactic miniframe for biopsy of intracranial lesions in conjunction with preoperative magnetic resonance imaging. Multiple trajectories and angles have been used with precision and safety.

Conclusion

The AXiiiS stereotactic miniframe is a feasible, safe, and disposable platform for multitrajectory intracranial biopsies. Compared with existing platforms, this novel device provides a more stable base and wider limits of trajectory angles with comparable accuracy and precision.